The neutron detection efficiency measurements of a 10 inch NE213 scientillator

dc.contributor.authorAbdulla Abdulaziz Al-Sunaidi
dc.date1992
dc.date.accessioned2022-05-18T04:04:04Z
dc.date.available2022-05-18T04:04:04Z
dc.degree.departmentCollege of Sciences
dc.degree.grantorKing Fahad for Petrolem University
dc.description.abstractA 10 inch NE213 scintillator in direct coupling to a 5 inch photomultiplier was modified by incorporating a light guide between the scintillation cell and the photomultiplier. Three tests were carried out to see the effect of the light guide on the performance of the detector. Results of the first test, which agreed with published data, showed an improvement in the detector energy resolution of about 50%. In time resolution test, an improvement of about 30% was achieved. Finally, no noticeable change in the neutron-gamma discrimination at operating biases was recorded. After the detector, the response function and the neutron detection efficiency of the modified detector were studied utilizing the Time-of-Flight technique and using a 252Cf fission neutron source.
dc.identifier.other4940
dc.identifier.urihttps://drepo.sdl.edu.sa/handle/20.500.14154/620
dc.language.isoen
dc.publisherSaudi Digital Library
dc.thesis.levelMaster
dc.thesis.sourceKing Fahad for Petrolem University
dc.titleThe neutron detection efficiency measurements of a 10 inch NE213 scientillator
dc.typeThesis